The Panametrics- NDT™ 25MX PLUS is an ultrasonic precision thickness gage that
provides cost-effective solutions in applications where the opposite side of the
test material is difficult or impossible to reach. Multipoint and single
thickness measurements can be made on a wide variety of materials including
metal, plastic, rubber, glass, fiberglass, and composites.
Features
- Single or multipoint thickness measurements
- Connects up to 8 transducers
- User selectable: number of channels, channel sequencing, and channel rate
- Uses contact, delay line, and immersion transducers
- Transducer
frequency range 2 - 30 MHz
- Wide thickness range, from 0.003 to 20 inches
(0.080 mm - 500 mm) depending on application
- Resolution up to 0.0001"
(0.001 mm)
- A-scan display for waveform verification
- Application
Auto-Recall with stored standard and custom setups
- Alphanumeric file-based
datalogger stores 18,000 thickness readings or 1,750 waveforms
- Handheld,
intuitive operation, long battery life
Mx-8 Multiplexer
Number of Channels: 1 to 8, user selectable
Channel Switching Rate: 1, 2, 4, 8, 16, or 20 Hz, user
selectable
Size: 3.2" x 5.9" x 1.25"(149.9 mm x 81.3 mm x 31.8 mm)
Weight: 7.7oz. (.22 kg)
Multipoint Thickness Measurements
Connected to the
optional MX-8 multiplexer, the 25MX PLUS has the ability to collect and display
thickness measurements from up to 8 separate transducers.
Special software
on the 25MX PLUS
allows the user to select:
- Number of channels (1-8)
- Mux Mode
- Off
-
Single Channel (manual switching)
-
Single
Sequence (switch through selected channels on command)
-
Sequence
(continuous channel switching)
- Channel Rate (1, 2, 4, 8, 16 or 20 Hz;
thickness output and channel number)
Single Thickness Measurements
The
25MX PLUS operates as a 'standard' precision thickness gage when not connected
to the MX-8 Multiplexer.
TECHNICAL Specifications
Measurements
Mode 1: |
Time interval between excitation pulse and
first backwall echo, using contact transducers. |
Mode 2: |
Time interval between the first interface
echo after the excitation pulse and the first backwall echo, using delay line
or
immersion transducers. |
Mode 3: |
Time interval between successive backwall
echoes following the first interface echo after the excitation pulse, using
delay line or immersion transducers. |
Thickness Measurement Range:
Steel: 0.006 - 20.000 inches
(0.150 - 508.00 mm)
Plastic: 0.003 - 2.0000 inches
(0.080 -50.00 mm)
Thickness range depends on material, transducer, surface condition,
temperature, and set up selected.
Material Velocity
Range:
0.02000 - 0.55110 inch/ µS
(0.5080 - 13.9979 mm/ µS)
Resolution: Keypad Selectable:
LOW: |
0.01" |
0.1 mm |
STANDARD: |
0.001" |
0.01 mm |
HIGH: |
0.0001" |
0.001 mm |
Power
Battery: 6 V Rechargeable NiCad battery pack, or
field-replaceable alkaline AA batteries
Battery Life: 25 hours in normal measurement mode with backlight off
Fast-Charger: Two-hour Fast Charger with universal voltage
Auto Power Off: On or off
General
Transducer Frequency
Range:
2-30 MHz (-3 dB)
Operating
Temperature:
-10 °C to +50 °C (+14 ° to 122 °F)
Keypad: Sealed color-coded keypad with tactile and
audible feedback
Case: Impact-resistant and
Water-resistant, gasketed Lexan ® case with sealed connectors. Environmental
IP-65 compliant
Metric/English Units
Min/Max Mode
Two
Alarm Modes: Programmable Hi-Low Alarm with audible and visual
indicators.
1) Standard Hi-Low
2) Previous thickness
Two Differential Modes:
1) Thickness
difference between actual measurement and reference value.
2) Percent
difference between actual measurement and reference value.
Application Auto-Recall: Automatically adjusts internal
parameters and zero offset for a wide variety of transducers.
Stored
Standard Setups: 25 stored transducer setups to allow fast, easy
calibration for our standard transducers.
Stored Custom
Setups: Up to 35 stored custom transducer setups for best performance
in special applications.
Size: 9.375 x 5.45 x 1.5" (238 x 138 x 38 mm)
Weight: 2.1 lbs. (0.95 kg)
INTERNAL DATALOGGER
Datalogger and RS-232: Identifies,
stores, recalls, clears, and transmits thickness readings, waveform images, and
gage setup information via the RS-232 Serial Port. Baud Rate, Word Length, Stop
Bits, and Parity are adjustable from the keypad.
Maximum # of
Stored Values:
Standard: |
18,000 thickness readings or 1,750
waveforms with thickness |
Upgrade 1: |
36,000 thickness readings or 3,400
waveforms with thickness |
Upgrade 2: |
54,000 thickness readings or 5,100
waveforms with thickness |
Location
Codes: |
8-character file name plus 16-character
alphanumeric location code input. Multiple comments per location. |
File Structures: |
7 standard or custom application-specific
file structures |
Reports: |
On-gage reporting of: Summary with
statistics, Min/Max with locations, and File Comparison. On-screen comparison
of
current and previous readings. |
DISPLAY
Display: Liquid Crystal Display
with backlight. Contrast keypad adjustable. Display area 4.0 x 3.39 inch (102
x
86 mm)
Backlight, Zoom Mode, Freeze Mode and Hold/Blank
Rectification: RF, half wave positive or negative, and full
wave
Waveform Display Range and Delay Control
Standard inclusions
Model 25MX PLUS Digital Ultrasonic
Thickness Gage, AC or Battery Operation, 50-60 Hz, with Internal Alphanumeric
Datalogger. Including: Universal Quick Charger/AC Adapter Transducer Cable, Test
Block, Couplant, RS-232 I/O Cable, Carrying Case, Instruction Manual and Two
Year Limited Warranty
Optional Accessories
MX8 |
MX-8 Multiplexer |
PLUS/RPC |
Rubber Protective Pouch with Neck Strap |
36DLP/SPC/KIT |
Protective Pouch with Neck Strap |
WIN25DLPLUS |
Interface Program |
25DLP/EW |
Extended Limited Warranty, 3rd Year |
2214E |
5-Step Test Block, 1018 steel, English
Units: .100", .200", .300", .400", .500"; metric available |
2213E |
5-Step Test Block, Aluminum, English Units:
.100", .200", .300", .400", .500"; metric available |
6DLPLUS/HDC |
Heavy Duty Shipping Case |
Please contact us for other accessories available with
the Model 25MX PLUS.
*All specifications are subject to change without notice.
Downloads